美国Cannon 凯能Cannon-Ubbelohde稀释型玻璃管粘度计稀释型乌氏粘度计简介:
The Cannon-Ubbelohde Dilution Viscometer measures intrinsic viscosity of transparent Newtonian liquids according to ASTM D445 and ISO 3104. Minimum sample volume is 8 mL (can be diluted to 40 mL). Minimum bath depth is 305 mm (12 in). The CANNON CT-518 bath is recommended. (CT-500 and CT-600 baths are not deep enough.) Viscometer constant is the same at all temperatures. Both calibrated and uncalibrated viscometers are available. Calibrated viscometers include a certificate of calibration and instruction sheet. Holders are sold separately.
根据ASTM D 445和ISO 3104的要求,Cannon-Ubbelohde稀释型乌氏粘度计测量透明牛顿液体的特性粘度。最小样品体积为8毫升(可稀释至40毫升)。最小沐浴深度为305毫米(12英寸)。推荐用 CANNON ct-518水浴。(CT-500和CT-600浴缸不够深。)粘度计常数在任何温度下都是相同的。校准和未校准粘度计都是可用的。校准粘度计包括校准证书和说明书。
美国Cannon凯能Cannon-Ubbelohde稀释型玻璃管粘度计稀释型乌氏粘度计技术参数:
CANNON-Ubbelohde Dilution Viscometer
Catalog Number | Size | Approximate Constant (mm²/s²) | Viscosity Range (mm²/s)† |
Uncalibrated | Calibrated | Low | High |
9722-L50 | 9722-M50 | 25 | 0.002 | 0.5 | 2.0 |
9722-L53 | 9722-M53 | 50 | 0.004 | 0.8 | 4.0 |
9722-L56 | 9722-M56 | 75 | 0.008 | 1.6 | 8.0 |
9722-L59 | 9722-M59 | 100 | 0.015 | 3.0 | 15 |
9722-L62 | 9722-M62 | 150 | 0.035 | 7.0 | 35 |
9722-L65 | 9722-M65 | 200 | 0.1 | 20 | 100 |
9722-L68 | 9722-M68 | 300 | 0.25 | 50 | 250 |
9722-L71 | 9722-M71 | 350 | 0.5 | 100 | 500 |
9722-L74 | 9722-M74 | 400 | 1.2 | 240 | 1200 |
9722-L77 | 9722-M77 | 450 | 2.5 | 500 | 2500 |
9722-L80 | 9722-M80 | 500 | 8.0 | 1600 | 8000 |
9722-L83 | 9722-M83 | 600 | 20 | 4000 | 20000 |
† Viscosity ranges in mm²/s represent flow times of 250 to 1000 seconds for size 25
and 200 to 1000 seconds for all other sizes
美国Cannon 凯能Cannon-Ubbelohde 稀释型乌氏粘度计 方法介绍:
美国Cannon玻璃毛细粘度管测试方法:ASTM D445, ASTM D446, ASTM D2170方法,透明和不透明液体运动粘度测定法 (包括计算动力粘度),玻璃毛细管运动粘度计的规格及使用说明,沥青运动粘度的试验方法
原理:本法用以测定透明与不透明产品的粘度,它是在已知的温度下,利用液体本身的重力流动性(gravity-flow)来测定一定体积的液体通过经校正后的毛细玻璃管粘度计所需的时间。根据所测出的动力粘度可计算出公制粘度或粘度(dynamic viscosity)。
方法概要:取一定体积的液体在严格的温度与固定的液面高度的控制下,使其流经毛细管粘度计而计算其流经时间。根据流经时间与粘度计的校正常数的乘积即可得动力粘度。
所需仪器:
1. 粘度计: 校正过后的玻璃毛细管,其所测得的粘度能符合于所要求的复验准确度再试准确度。
2. 粘度计支架: 能将粘度计在恒温槽中架持稳固,且能使其垂直竖立。
3. 粘度计调温槽: 内盛装透明液体,能使样品于粘度计内较液面低20mm以下,或高于槽底20mm以上,能保持槽内温度的变动于15至100℃间低于±0.02℃的变化,在此温度范围以外,亦不超过±0.05℃的温度变化。
4. 测温器: 检验动力粘度计的温度计内盛有适当的液体,使之能符合温度范围,温度计使用前均需校正,校正后准确度在±0.02℃,若其它测温器有上述的精确度时亦可采用。
5. 定时器: 秒表或弹簧装置的定时器均可使用,但其刻度至少在0.1秒,其测定200秒至900秒的误差在±0.07%以内。